Reliability of the AMS inflatable prostheses in 444 consecutive cases.

作者信息

Scarzella G I

机构信息

George Washington University, Washington, DC.

出版信息

Arch Ital Urol Androl. 1993 Dec;65(6):661-4.

PMID:8312948
Abstract

Between February, 1981, and September, 1990, 444 men with erectile dysfunction were implanted with AMS inflatable penile prostheses. Etiology was vascular for 62% and diabetic for 22%. Ninety-three percent had a functional prosthesis at last contact. Overall, 74% of the originally-implanted cylinders were functional at last contact with no revisions; an additional 3% required one or more revisions, though no replacement, to maintain function. Of the sub-group receiving PPS cylinders as the first implant under my care, 35% have required no revision; along with 66% implanted with the thicker-walled PPT; 85% with the PND, designed to reduce stress on the tunica albuginea and prevent aneurysm formation; and 90% with the CX cylinders. This greatly improved reliability can be ascribed to (1) continued improvements by the manufacturer, (2) use of a single prosthesis type leading to improved surgical proficiency, and (3) a modified revision technique when a fluid leak has occurred, making it frequently unnecessary to replace the entire device.

摘要

在1981年2月至1990年9月期间,444例勃起功能障碍男性接受了美国医疗系统公司(AMS)可膨胀阴茎假体植入。病因中血管性的占62%,糖尿病性的占22%。在最后一次随访时,93%的患者假体功能良好。总体而言,最初植入的圆柱体在最后一次随访时74%功能良好且无需翻修;另有3%虽无需更换但需要一次或多次翻修以维持功能。在我负责的首次植入PPS圆柱体的亚组中,35%无需翻修;首次植入厚壁PPT圆柱体的患者中这一比例为66%;首次植入旨在减轻白膜压力并预防动脉瘤形成的PND圆柱体的患者中这一比例为85%;首次植入CX圆柱体的患者中这一比例为90%。可靠性的大幅提高可归因于:(1)制造商的持续改进;(2)使用单一类型的假体提高了手术熟练程度;(3)出现液体泄漏时采用改良的翻修技术,使得常常无需更换整个装置。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验