Losordo D W, Pastore J O, Coletta D, Kenny D, Isner J M
Department of Medicine (Cardiology), St. Elizabeth's Hospital, Tufts University School of Medicine, Boston, MA 02135.
Am Heart J. 1993 Jul;126(1):168-76. doi: 10.1016/s0002-8703(07)80025-1.
The objective of this study was to determine the validity of estimation of regurgitant volume by visual assessment of color flow Doppler display. An experimental apparatus was designed that is capable of ejecting precise volumes of echogenic material from one chamber to another under continuous color flow Doppler monitoring. The velocity of flow was altered independently by changing either the size of the orifice through which flow occurred or the ejection rate. In this manner the differential effects of volume and velocity on the color flow Doppler image could be examined. The maximum area encompassed by the color flow Doppler pattern for each ejection was planimetered by using commercially available on-line software. In addition the reviewer in each case applied a subjective grade to the appearance of the color flow jet (1+ to 4+). Comparison was then made of the color flow Doppler appearance of equal volumes flowing at different velocities and of different volumes flowing at different velocities. In the initial series a solution of agitated hetasarch was used. When equal volumes were imaged at different velocities the higher-velocity jet appeared larger, both subjectively (3+ vs 1+) and by measuring the area encompassed in the Doppler flow profile (40.3 +/- 1.8 vs 22.0 +/- 1.4 cm2, p = 0.0001). Furthermore, when different volumes were imaged at different velocities, the smaller volume (3 ml vs 6 ml) appeared larger when it was flowing at higher velocity (3+ vs 2+, 40.3 +/- 1.8 vs 32.4 +/- 1.3 cm2, p = 0.0006). These experiments were repeated with blood, confirming the results of the initial study.(ABSTRACT TRUNCATED AT 250 WORDS)
本研究的目的是确定通过彩色多普勒血流显像的视觉评估来估计反流容积的有效性。设计了一种实验装置,该装置能够在连续彩色多普勒监测下将精确体积的回声物质从一个腔室喷射到另一个腔室。通过改变血流通过的孔口大小或喷射速率来独立改变血流速度。通过这种方式,可以研究容积和速度对彩色多普勒图像的不同影响。使用市售的在线软件对每次喷射的彩色多普勒血流图所包含的最大面积进行面积测量。此外,在每种情况下,评估者对彩色血流束的外观进行主观分级(1+至4+)。然后比较不同速度下相同容积血流以及不同速度下不同容积血流的彩色多普勒表现。在最初的系列实验中使用了搅拌过的贺斯溶液。当相同容积以不同速度成像时,主观上(3+对1+)以及通过测量多普勒血流剖面图中所包含的面积(40.3±1.8对22.0±1.4cm²,p=0.0001),较高速度的血流束看起来更大。此外,当不同容积以不同速度成像时,较小容积(3ml对6ml)在较高速度下流动时(3+对2+,40.3±1.8对32.4±1.3cm²,p=0.0006)看起来更大。用血液重复了这些实验,证实了最初研究的结果。(摘要截断于250字)