Ishigooka M, Hashimoto T, Izumiya K, Katoh T, Yaguchi H, Nakada T, Handa Y, Hoshimiya N
Department of Urology, Yamagata University, School of Medicine, Japan.
Front Med Biol Eng. 1993;5(1):1-10.
Electrical pelvic floor stimulation employing a portable functional electrical stimulation system with percutaneously indwelling electrodes was carried out to improve detrusor urinary incontinence. Cyclic stimulation using negative going pulse trains of 20 Hz was applied 3 to 6 times daily to the bilateral pudendal nerves distributing to the pelvic floor muscles for the purpose of strengthening these muscles, including the urethral sphincter, and simultaneously, suppressing detrusor overactivity and increasing cystometric capacity. Electrical training for 4-8 weeks resulted in an improvement of urinary incontinence in five of six patients. In two of six cases incontinence had subjectively disappeared. Urodynamic investigations demonstrated an increase in detrusor reflex threshold and less tendency for abortive detrusor contraction. No apparent complications were encountered during these periods. This procedure appears to be efficient for the management of patients with detrusor incontinence who respond poorly to conservative therapies.
采用带有经皮留置电极的便携式功能性电刺激系统进行盆底电刺激,以改善逼尿肌性尿失禁。使用20Hz的负向脉冲串进行周期性刺激,每天对分布于盆底肌肉的双侧阴部神经施加3至6次,目的是加强包括尿道括约肌在内的这些肌肉,同时抑制逼尿肌过度活动并增加膀胱容量测定值。4至8周的电刺激训练使6例患者中的5例尿失禁情况得到改善。6例中有2例患者的失禁症状主观上消失。尿动力学检查显示逼尿肌反射阈值增加,逼尿肌收缩中断的趋势降低。在此期间未出现明显并发症。对于对保守治疗反应不佳的逼尿肌性尿失禁患者,该方法似乎有效。