Amoore J N
Department of Medical Physics and Medical Engineering, Royal Infirmary of Edinburgh, UK.
J Med Eng Technol. 1993 Jan-Feb;17(1):25-31. doi: 10.3109/03091909309009980.
The assessment of the functional performance of non-invasive blood pressure (NIBP) monitors using a commercially available NIBP analyser (Dynatech CuffLink) was evaluated. The CuffLink includes facilities for performing a static pressure calibration, and safety cut-out pressure, leak rate and rapid deflate time measurements. Cuff pressure oscillations simulating a range of blood pressures at a range of heart rates enable dynamic tests of NIBP monitors to be performed. By reducing the magnitude of the cuff pressure oscillations the sensitivity of a monitor to weak pulsations can be determined. The CuffLink does not provide facilities for assessing the ability of the monitor to reject movement artifact. The CuffLink tests were performed on 10 different models of NIBP monitor. The results demonstrate the variable nature of systolic, diastolic and mean blood pressure readings from NIBP monitors (even under stable laboratory conditions), reinforcing the importance of not relying on individual readings, but rather to consider the average of a few successive readings. Normative data for two models of NIBP monitor are provided. Applications of the CuffLink are discussed with reference to a protocol for NIBP monitor performance checks.
使用市售的无创血压(NIBP)分析仪(Dynatech CuffLink)对NIBP监测仪的功能性能进行了评估。CuffLink具备进行静压校准、安全切断压力、泄漏率和快速放气时间测量的功能。模拟一系列心率下的一系列血压的袖带压力振荡能够对NIBP监测仪进行动态测试。通过减小袖带压力振荡的幅度,可以确定监测仪对微弱脉动的敏感度。CuffLink没有提供评估监测仪抑制运动伪影能力的功能。在10种不同型号的NIBP监测仪上进行了CuffLink测试。结果表明,NIBP监测仪的收缩压、舒张压和平均血压读数具有变异性(即使在稳定的实验室条件下),这进一步强调了不要依赖单个读数,而应考虑连续几次读数平均值的重要性。提供了两种型号NIBP监测仪的规范数据。参照NIBP监测仪性能检查方案讨论了CuffLink的应用。