Hanna K D, Hayward J M, Hagen K B, Simon G, Parel J M, Waring G O
Opthalmology Department, Hôpital Hotel-Dieu, Paris, France.
Arch Ophthalmol. 1993 Jul;111(7):998-1004. doi: 10.1001/archopht.1993.01090070118030.
We describe a new system for performing transverse arcuate corneal incisions for the correction of astigmatism. The instrument consists of a conical housing with suction and/or friction fixation to the limbus and two micrometric diamond knives mounted on a carrier that can be rotated by a gear mechanism. The depth, length, and location of the incisions can be predetermined. Light microscopy of donor eye corneas documented the achieved incision depth. One sweep at normal speed produced an incision depth 28% less deep than the preset depth. One sweep at low speed achieved depths of approximately 100% of the knife setting. Variability from one incision to another in the same cornea was related to blade sharpness. We report successful use of the instrument in five consecutive eyes with naturally occurring astigmatism.
我们描述了一种用于进行横向弧形角膜切口以矫正散光的新系统。该仪器由一个带有用于固定在角膜缘的吸力和/或摩擦力的圆锥形外壳,以及安装在一个可通过齿轮机构旋转的载体上的两把微米级金刚石刀组成。切口的深度、长度和位置可以预先确定。对供体眼角膜的光学显微镜检查记录了所达到的切口深度。以正常速度进行一次扫描产生的切口深度比预设深度浅28%。以低速进行一次扫描可达到刀设置深度的约100%。同一角膜中一个切口与另一个切口之间的差异与刀片的锋利程度有关。我们报告了该仪器在连续五只患有自然散光的眼睛中的成功应用。