一种通过压电薄膜测量鸟类胚胎和幼雏心率的简单、无创系统。

Simple, noninvasive system for measuring the heart rate of avian embryos and hatchlings by means of a piezoelectric film.

作者信息

Tazawa H, Hashimoto Y, Takami M, Yufu Y, Whittow G C

机构信息

Department of Electrical & Electronic Engineering, Muroran Institute of Technology, Japan.

出版信息

Med Biol Eng Comput. 1993 Mar;31(2):129-34. doi: 10.1007/BF02446670.

Abstract

Using a flexible piezoelectric film, the authors developed a simple system to determine noninvasively the heart rate of chicken embryos and hatchlings. The film was piezoelectric polyvinylidene fluoride (PVDF), which is sensitive enough to detect cardiogenic ballistic movements of the egg (ballistocardiogram (BCG)) and precordial movements of the hatchling attributable to cardiac contractions (apexcardiogram (ACG)). The BCG could be detected, during the second half of incubation, by placing the egg on the PVDF film on a soft substrate. The detected signal was found to be a measure of movement velocity. The ACG could be measured when the hatchling's chest wall made contact with the PVDF film installed in a box in which the hatchling was confined. The heart rate was counted from the lag time of autocorrelation calculated for a certain time segment (e.g. 2 s) of the BCG and ACG recordings.

摘要

作者使用一种柔性压电薄膜开发了一个简单系统,用于非侵入性地测定鸡胚胎和雏鸡的心率。该薄膜为聚偏二氟乙烯(PVDF)压电薄膜,其灵敏度足以检测鸡蛋的心源性弹道运动(心冲击图(BCG))以及雏鸡因心脏收缩而产生的心前区运动(心尖搏动图(ACG))。在孵化后半期,将鸡蛋放置在柔软基质上的PVDF薄膜上,即可检测到BCG。检测到的信号是运动速度的一种度量。当雏鸡的胸壁与安装在雏鸡所在盒子中的PVDF薄膜接触时,即可测量ACG。通过计算BCG和ACG记录特定时间段(例如2秒)的自相关滞后时间来计算心率。