MettingVanRijn A C, Kuiper A P, Linnenbank A C, Grimbergen C A
Medical Physics & Informatics Department, University of Amsterdam, Faculty of Medicine, The Netherlands.
IEEE Trans Biomed Eng. 1993 Mar;40(3):302-8. doi: 10.1109/10.216416.
A design for patient isolation in 64-channel ECG recordings is presented. Small dimensions of the isolated section and the use of an optical fiber as the only connection between the isolated section and the grounded section of the measurement system ensured a minimal capacitance between the patient and the environment. The consistent low-power design of the isolated section resulted in a power consumption of 210 mW, which enabled a 10 h continuous operating time of the battery powered isolated section. The system handles 64 signals with a dynamic range of 75 dB. Analog to digital conversion is performed in the isolated section with a sample rate of 1 kHz per channel. The receiver interfaces to a commercially available DMA board for a standard personal computer.