Laser Doppler flowmetry for estimation of bone blood flow: studies of reproducibility and correlation with microsphere technique.

作者信息

Lausten G S, Kiaer T, Dahl B

机构信息

Department of Orthopedics, Rigshospitalet, University of Copenhagen, Denmark.

出版信息

J Orthop Res. 1993 Jul;11(4):573-80. doi: 10.1002/jor.1100110412.

Abstract

The blood flow in the femoral condyles of six pigs was measured by the microsphere technique and by laser Doppler flowmetry (LDF) before and after clamping of the external iliac artery. A significant decrease in the bone blood flow was found with both microsphere and LDF measurements, and a high correlation was found between the two methods. After release of the arterial clamp, the LDF values rapidly returned to approximately the same values as before clamping. In a second part of the study, the LDF output signal was studied in the intertrochanteric area of eight patients. The reproducibility of two LDF measurements at the same location was within 15%. The temporal variability of the LDF signal during 10 min of recording showed only slight variations. It is concluded that LDF is relative in nature and that the mean LDF output signal is suitable for monitoring dynamic responses of the bone blood flow to various stimuli. Several consecutive measurements have to be performed if the blood flow at different locations of a specific bone are to be compared with each other.

摘要

采用微球技术和激光多普勒血流仪(LDF)测量6只猪在髂外动脉夹闭前后股骨髁的血流情况。微球测量法和LDF测量法均发现骨血流显著减少,且两种方法之间具有高度相关性。松开动脉夹后,LDF值迅速恢复到与夹闭前大致相同的值。在研究的第二部分,对8例患者转子间区域的LDF输出信号进行了研究。在同一位置进行的两次LDF测量的重复性在15%以内。记录10分钟期间LDF信号的时间变异性仅显示出轻微变化。得出的结论是,LDF本质上是相对的,平均LDF输出信号适用于监测骨血流对各种刺激的动态反应。如果要比较特定骨骼不同位置的血流,则必须进行多次连续测量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验