Suppr超能文献

肠系膜上动脉的双功超声检查:观察者间的变异性

Duplex ultrasonography of superior mesenteric artery: interobserver variability.

作者信息

Perko M J, Just S

机构信息

Department of Vascular Surgery, Rigshospitalet, Copenhagen, Denmark.

出版信息

J Ultrasound Med. 1993 May;12(5):259-63. doi: 10.7863/jum.1993.12.5.259.

Abstract

The aim of this study was to evaluate the interobserver variability during duplex ultrasonographic investigation of the SMA. Eleven healthy volunteers were examined twice under the same conditions by two independent observers. Three basic Doppler wave parameters were assessed for the SMA: pV, edV, and rT. The tD was recorded for SMA and abdominal aorta (AotD) with B-mode imaging. The interobserver variability was very low for all parameters: pV = 13.5 cm/s; edV = 3.4 cm/s; rT = 0.01 s; tD = 0.3 mm; AotD = 0.5 mm, with corresponding 95% confidence intervals as follows: pV = 19 cm/s; edV = 4.7 cm/s; rT = 0.015 s; tD = 0.4 mm; and AotD = 0.75 mm. Coefficients of variation also showed very low dispersion for results obtained. Under the appropriate standard conditions for duplex sonographic examination of the mesenteric artery, assessments of basic Doppler wave parameters do not differ significantly between independent investigators. Reliable evaluation of B-mode distances can be obtained by replication of measurements.

摘要

本研究的目的是评估在肠系膜上动脉(SMA)的双功超声检查过程中观察者间的变异性。11名健康志愿者在相同条件下由两名独立观察者进行了两次检查。对SMA评估了三个基本多普勒波参数:峰值流速(pV)、舒张末期流速(edV)和阻力指数(rT)。通过B型成像记录了SMA和腹主动脉的管径(tD和腹主动脉管径[AotD])。所有参数的观察者间变异性都非常低:pV = 13.5 cm/s;edV = 3.4 cm/s;rT = 0.01 s;tD = 0.3 mm;AotD = 0.5 mm,相应的95%置信区间如下:pV = 19 cm/s;edV = 4.7 cm/s;rT = 0.015 s;tD = 0.4 mm;AotD = 0.75 mm。变异系数也显示所获得结果的离散度非常低。在肠系膜动脉双功超声检查的适当标准条件下,独立研究者对基本多普勒波参数的评估没有显著差异。通过重复测量可以获得对B型图像距离的可靠评估。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验