Weisskoff R M, Cohen M S, Rzedzian R R
Massachusetts General Hospital, Department of Radiology, Charlestown 02129.
Magn Reson Med. 1993 Jun;29(6):796-803. doi: 10.1002/mrm.1910290612.
We demonstrate that whole-body, single-shot imaging is practical for imaging out of the central plane, including oblique axes. The technique is illustrated by images of the heart in the cardiac long- and short-axis and by coronal images of the brain. Secondary gradients can produce additional image distortion and ghosting in these images. These artifacts are a direct consequence, predictable by Maxwell's equations, of the large gradients used in echo planar imaging. We show that these effects are, in general, made smaller by working at a high magnetic field.