Suppr超能文献

Development of narrow-band lamps for testing and clinical use.

作者信息

Alsins J, Claesson S, Fischer T, Juhlin L

出版信息

Strahlentherapie. 1977 Jan;153(1):61-4.

PMID:835148
Abstract

A compact narrow band lamp for high intensity irradiation has been developed. It consists of a high-pressure mercury lamp, a shutter and water cooled filters for spectral isolation of the light. A mixing device permits simultaneous irradiation with two different wavelenghts by using two lamps. The spectral distribution of the light within the bands is given for seven filter combinations. The construction and calibration of a power meter is also described.