使用罗氏Cobas Core免疫化学分析仪对甲胎蛋白异质体进行快速、自动化酶免疫分析。

Rapid, automated enzyme immunoassay for MCA with the Cobas Core Roche immunochemistry analyser.

作者信息

Henne V, Maurer A

机构信息

Roche Diagnostic Systems, Division of F. Hoffmann-La Roche Ltd., Basle, Switzerland.

出版信息

Int J Biol Markers. 1993 Apr-Jun;8(2):133-7. doi: 10.1177/172460089300800211.

Abstract

The breast cancer marker Mucin-like Carcinoma-associated Antigen (MCA) was evaluated in a version that has been adapted to a new enzyme immunoanalyser, the Cobas Core. The two-step sandwich assay has a short total assay time of less than 60 minutes. Standard curves are stable for at least 2 weeks. The coefficients of variation for the inter-assay precision ranged between 3.6% and 7.8%, but were mainly below 5%. Single determinations of patient values from different runs correlated also very closely (R = 0.99). The dilution linearity for a set of 14 sera exhibited recovery rates mostly ranging from 90% to 110%. From 165 blood donors investigated, more than 99% had values below the cut-off of 11 U/ml. As demonstrated for pathological sera, the automated and manual versions showed a high degree of accordance (R = 0.98). The investigation presents convincing evidence that the Cobas Core MCA EIA running on a Cobas Core analyser guarantees a high analytical reliability and reproducibility.

摘要

在已适配新型酶免疫分析仪Cobas Core的版本中,对乳腺癌标志物黏蛋白样癌相关抗原(MCA)进行了评估。两步夹心测定法的总测定时间较短,不到60分钟。标准曲线至少2周内稳定。批间精密度的变异系数在3.6%至7.8%之间,但大多低于5%。不同批次患者值的单次测定相关性也非常紧密(R = 0.99)。一组14份血清的稀释线性显示回收率大多在90%至110%之间。在调查的165名献血者中,超过99%的人的值低于11 U/ml的临界值。正如病理血清所显示的那样,自动版和手动版显示出高度一致性(R = 0.98)。该研究提供了令人信服的证据,表明在Cobas Core分析仪上运行的Cobas Core MCA酶免疫测定法保证了高分析可靠性和可重复性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验