Karger A E, Weiss R, Gesteland R F
Department of Human Genetics, University of Utah, Salt Lake City.
Anal Chem. 1993 Jul 1;65(13):1785-93. doi: 10.1021/ac00061a024.
A cryogenically cooled charge-coupled device (CCD) camera equipped with an area CCD array is used in a line scanning system for low-light-level imaging of chemiluminescent DNA sequencing blots. Operating the CCD camera in time-delayed integration (TDI) mode results in continuous data acquisition independent of the length of the CCD array. Scanning is possible with a resolution of 1.4 line pairs/mm at the 50% level of the modulation transfer function. High-sensitivity, low-light-level scanning of chemiluminescent direct-transfer electrophoresis (DTE) DNA sequencing blots is shown. The detection of DNA fragments on the blot involves DNA-DNA hybridization with oligonucleotide-alkaline phosphatase conjugate and 1,2-dioxetane-based chemiluminescence. The width of the scan allows the recording of up to four sequencing reactions (16 lanes) on one scan. The scan speed of 52 cm/h used for the sequencing blots corresponds to a data acquisition rate of 384 pixels/s. The chemiluminescence detection limit on the scanned images is 3.9 x 10(-18) mol of plasmid DNA. A conditional median filter is described to remove spikes caused by cosmic ray events from the CCD images.
一种配备面阵电荷耦合器件(CCD)阵列的低温冷却CCD相机用于线扫描系统,对化学发光DNA测序印迹进行低光水平成像。以延时积分(TDI)模式操作CCD相机可实现独立于CCD阵列长度的连续数据采集。在调制传递函数的50%水平下,扫描分辨率可达1.4线对/毫米。展示了对化学发光直接转移电泳(DTE)DNA测序印迹的高灵敏度、低光水平扫描。印迹上DNA片段的检测涉及DNA与寡核苷酸 - 碱性磷酸酶共轭物的DNA - DNA杂交以及基于1,2 - 二氧杂环丁烷的化学发光。扫描宽度允许在一次扫描中记录多达四个测序反应(16个泳道)。用于测序印迹的52厘米/小时的扫描速度对应于384像素/秒的数据采集速率。扫描图像上的化学发光检测限为3.9×10⁻¹⁸摩尔的质粒DNA。描述了一种条件中值滤波器,用于从CCD图像中去除由宇宙射线事件引起的尖峰。