Suppr超能文献

角膜赫斯堡-巴伦真空环钻造成的内皮损伤。

Endothelial damage by the corneal Hessburg-Barron vacuum trephine.

作者信息

Legeais J M, Parel J M, Simon G, Ren Q, Denham D

机构信息

Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami School of Medicine, FL 33101.

出版信息

Refract Corneal Surg. 1993 Jul-Aug;9(4):255-8.

PMID:8398970
Abstract

BACKGROUND

Tissue damage in corneal transplantation procedures was previously assessed in comparative pilot studies using vacuum-assisted metal blade trephines and a noncontact hydrogen fluoride laser trephine system. With the laser, endothelial damage was less than 20 x 30 microns at the edge of the posterior corneal perforation.

METHODS

In this study, a second generation vacuum-assisted Hessburg-Barron trephine was evaluated in full-thickness penetrating trephination performed in 16 rabbit eyes and eight fresh human donor eyes.

RESULTS

We found a relatively large, 150-micrometer wide, annular zone of endothelial damage on the recipient cornea located 1.25 mm away from the trephined opening. Our study demonstrates that endothelial damage is linked to the instrument's vacuum fixation ring.

CONCLUSION

With this trephine, endothelial damage may be avoided by using partial-depth trephination, filling the anterior chamber with viscoelastic and completing the procedure using a metal blade or diamond knife.

摘要

背景

在先前的比较性初步研究中,使用真空辅助金属刀片环钻和非接触式氟化氢激光环钻系统对角膜移植手术中的组织损伤进行了评估。使用激光时,角膜后穿孔边缘的内皮损伤小于20×30微米。

方法

在本研究中,对16只兔眼和8只新鲜人供体眼进行全层穿透环钻时,评估了第二代真空辅助赫斯堡-巴伦环钻。

结果

我们在距环钻开口1.25毫米处的受体角膜上发现了一个相对较大的、宽150微米的内皮损伤环形区域。我们的研究表明,内皮损伤与器械的真空固定环有关。

结论

使用这种环钻时,可通过进行部分深度环钻、用粘弹性物质填充前房并使用金属刀片或钻石刀完成手术来避免内皮损伤。