Quality assurance for intraoperative transesophageal echocardiography monitoring: a report of 846 procedures.

作者信息

Rafferty T, LaMantia K R, Davis E, Phillips D, Harris S, Carter J, Ezekowitz M, McCloskey G, Godek H, Kraker P

机构信息

Department of Anesthesiology, Yale University School of Medicine, Yale-New Haven Hospital, Connecticut 06510.

出版信息

Anesth Analg. 1993 Feb;76(2):228-32. doi: 10.1213/00000539-199302000-00005.

Abstract

We evaluated our experience with 846 consecutive transesophageal echocardiography (TEE) intraoperative monitoring procedures performed between November 1989 and July 1991. TEE frequency was 36 +/- 11 per month (range 16-55) and represented 69.8% of cardiac valve surgery cases, 40.2% of coronary artery bypass graft cases, and 2.2% of total operative caseload. Major patient complications consisted of transient vocal cord paresis and ingestion of glutaraldehyde-disinfectant solution. Minor complications consisted of a chipped tooth (one case) and pharyngeal abrasions (three cases). The Quality Assurance (Q/A) Program evaluated both record keeping and quality of imaging, as judged by cardiologist echocardiographer reviewers. The percentage of completion for each Q/A indicator was as follows: medical record documentation, 88%; database form annotation, 94%; and provision of videotape recording, 91%. TEE database forms were analyzed further in terms of the percentage of fields completed. Completion scores were 73%. The following scoring system was utilized for videotape evaluation by the cardiologists: 1 = excellent; 2 = good; 3 = poor. The median grade for both two-dimensional echocardiography and color flow Doppler (CFD) examinations was 2. Poor quality images (grade 3) were present in 15.2% of two-dimensional echocardiography and 20.3% of color flow Doppler examinations, and disproportionately associated with 4/26 attendings. Supplemental audit of the cardiology reviewers performance demonstrated 569/846 videotapes showed no objective evidence of review. The cardiology reviewer forms of the remaining 277 videotapes were evaluated in terms of the percentage of fields completed. The completion score was 56%. These data suggest the need for formal Q/A for intraoperative TEE, both for anesthesiologists and reviewing cardiologists.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验