Sciscione A C, Manley J S, Pinizzotto M E, Colmorgen G H
Department of Obstetrics and Gynecology, Medical Center of Delaware, Christiana Hospital, Newark 19718.
Am J Perinatol. 1993 Jan;10(1):21-3. doi: 10.1055/s-2007-994693.
There have been few reported complications of intrauterine pressure monitoring. We present four cases of placental abruption following insertion of a disposable intrauterine pressure transducer system (INTRAN). All four patients had placental abruptions, which were evident soon after insertion of the disposable intrauterine pressure transducer system (all of our cases had predisposing factors or signs of placental abruption at presentation). Whether there was a preexisting placental abruption, which was caused, revealed, or worsened by insertion of the disposable intrauterine pressure transducer system is unclear. We encourage careful patient selection with gentle insertion of the system opposite the placental site.
关于宫内压力监测的并发症报告较少。我们报告了4例在插入一次性宫内压力换能器系统(INTRAN)后发生胎盘早剥的病例。所有4例患者均发生了胎盘早剥,在插入一次性宫内压力换能器系统后很快就表现明显(我们所有病例在就诊时均有胎盘早剥的诱发因素或迹象)。一次性宫内压力换能器系统的插入是否导致、揭示或加重了先前存在的胎盘早剥尚不清楚。我们鼓励谨慎选择患者,并在胎盘部位对侧轻柔插入该系统。