Pierce W S, Snyder A J, Rosenberg G, Weiss W J, Pae W E, Waldhausen J A
Department of Surgery, College of Medicine, Pennsylvania State University, Hershey 17033.
J Thorac Cardiovasc Surg. 1993 Mar;105(3):520-4.
An implantable, electrically powered pump that will provide tether-free circulatory support is being developed. The blood pump consists of a seamless polyurethane sac within a polysulfone case. Björk-Shiley Monostrut valves provide unidirectional flow. The blood sac is compressed by a pusher plate with a stroke of 1.9 cm and actuated by a brushless direct-current electric motor and motion translator. The current unit is completely sealed, and inductive coupling techniques provide the electrical energy. The system has an implantable electronic control system as well as a battery that provides 30 minutes of operation when the external coil is disconnected. During normal operation, however, the pump is powered by a portable battery pack or by house current. The unit can pump 8.5 L/min at physiologic pressures. Twenty-six animals have had circulatory support for a period of more than 1 week. The average period of pumping was 62 days; the longest was nearly 8 months. Experiments were terminated in 18 animals because of pump-related problems and in 8 because of biologically related problems. Studies to date are very encouraging and suggest that, with further refinement, a reliable 2-year assist pump that will have important clinical application can be developed.
一种可植入的电动泵正在研发中,它将提供无束缚的循环支持。该血泵由聚砜外壳内的无缝聚氨酯囊组成。Björk-Shiley Monostrut瓣膜提供单向血流。血囊由行程为1.9厘米的推板压缩,并由无刷直流电动机和运动转换器驱动。当前的装置完全密封,感应耦合技术提供电能。该系统有一个可植入的电子控制系统以及一个电池,当外部线圈断开连接时,电池可提供30分钟的运行时间。然而,在正常运行期间,泵由便携式电池组或市电供电。该装置在生理压力下每分钟可泵血8.5升。26只动物已获得超过1周的循环支持。平均泵血时间为62天;最长接近8个月。18只动物因与泵相关的问题终止实验,8只因生物学相关问题终止实验。迄今为止的研究非常令人鼓舞,表明经过进一步改进,可以开发出一种可靠的、具有重要临床应用价值的两年期辅助泵。