The effects of axial rotational alignment of the femoral component on knee stability and patellar tracking in total knee arthroplasty demonstrated on autopsy specimens.

作者信息

Anouchi Y S, Whiteside L A, Kaiser A D, Milliano M T

机构信息

DePaul Biomechanical Research Lab, St. Louis, MO 63044.

出版信息

Clin Orthop Relat Res. 1993 Feb(287):170-7.

PMID:8448937
Abstract

Four fresh-frozen anatomic knee specimens were tested for knee stability, patellar tracking, and patellofemoral contact points with the femoral component positioned in 5 degrees internal, 5 degrees external, or neutral axial rotational alignment of the femoral component referenced on the posterior femoral condyles. The externally rotated specimens had varus-valgus stability of the knee that was closest to the normal control. The internally rotated specimens shifted into valgus alignment with flexion. Patellar tracking also was closest to normal in the externally rotated specimens. Patellofemoral contact was more evenly distributed between the medial and lateral contact areas in the externally rotated specimens than in the internally rotated or in the neutral specimens. Internal rotation of the femoral component in the knee with perpendicular resection of the tibia causes undesirable changes in knee stability, patellar tracking, and patellofemoral contact points. Neutral positioning produces similar but less negative effects on knee stability and patellar kinematics. External rotation improves both patellar tracking and knee stability characteristics.

摘要

对四个新鲜冷冻的膝关节解剖标本进行了测试,以评估膝关节稳定性、髌骨轨迹以及股骨组件处于相对于股骨后髁的5度内旋、5度外旋或中立轴向旋转对线时的髌股接触点。外旋标本的膝关节内外翻稳定性最接近正常对照。内旋标本在屈曲时会向外翻对线偏移。外旋标本的髌骨轨迹也最接近正常。与内旋或中立标本相比,外旋标本的髌股接触在内外侧接触区域之间分布更均匀。在胫骨垂直截骨的膝关节中,股骨组件的内旋会导致膝关节稳定性、髌骨轨迹和髌股接触点出现不良变化。中立位对线对膝关节稳定性和髌骨运动学产生类似但负面影响较小。外旋可改善髌骨轨迹和膝关节稳定性特征。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验