Stetter J R, Sedlak J M, Blurton K F
J Chromatogr Sci. 1977 Mar-Apr;15(3-4):125-8. doi: 10.1093/chromsci/15.3.125.
An electrochemical detector for gas chromatographic analysis of hydrogen sulfide is described and its operating characteristics are discussed. The detector operation is based upon the measurement of the current when hydrogen sulfide is electrochemically oxidized at a diffusion electrode. The lower detectable limit was 3 X 10(-12) grams H2S, the precision was 0.5% and analysis was achieved within 2 minutes. Accuracy was limited principally by the preparation of calibration samples and the availability of standard reference gas mixtures.
描述了一种用于气相色谱分析硫化氢的电化学检测器,并讨论了其操作特性。该检测器的操作基于在扩散电极上对硫化氢进行电化学氧化时电流的测量。最低检测限为3×10⁻¹²克硫化氢,精密度为0.5%,且在2分钟内即可完成分析。准确度主要受校准样品的制备和标准参考气体混合物可用性的限制。