Abstract

In eight unique challenges mailed by The College of American Pathologists Comprehensive Chemistry Survey to participating laboratories within 3 years, results for direct-reacting bilirubin (DBIL) were highly variable among the 12 largest peer groups, and most of the results differed greatly from the values obtained by a referred method. Peer-group mean values for total bilirubin (TBIL) were in much better agreement with each other and with those obtained by the Reference Method for TBIL. From a review of the information on the assay of DBIL provided to us by the manufacturers, we conclude that among the major causes of the large variability and bias in DBIL assays are problems with calibration, lack of a serum blank measurement, inadequate concentrations of HCl in the reaction mixture, inappropriate use of bichromatic correction methods, and possibly the use of wetting agents or surfactants in the reagent. Within-group SDs were small and generally acceptable. The among-peer-group variability in DBIL values is attributable to bias, not imprecision. We recommend several simple changes that could improve the accuracy of DBIL determinations in clinical laboratories.

摘要

在美国病理学家学会综合化学调查机构在3年内寄给参与实验室的8项独特挑战中,12个最大的同类实验室组中直接反应胆红素(DBIL)的结果差异很大,并且大多数结果与参考方法获得的值有很大差异。总胆红素(TBIL)的同类组平均值彼此之间以及与通过TBIL参考方法获得的值之间的一致性要好得多。通过审查制造商提供给我们的关于DBIL检测的信息,我们得出结论,DBIL检测中存在较大变异性和偏差的主要原因包括校准问题、缺乏血清空白测量、反应混合物中HCl浓度不足、双色校正方法使用不当以及试剂中可能使用了湿润剂或表面活性剂。组内标准差较小且通常可以接受。DBIL值在同类组之间的变异性归因于偏差,而非不精密度。我们建议进行一些简单的改变,以提高临床实验室中DBIL测定的准确性。

相似文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验