A Marlex strut, which was designed to prevent efferent valve prolapse and peristomal herniation in the Kock continent urinary reservoir, eroded through the pouch wall, resulting in persistent bacteriuria and the inability to catheterize the pouch. The eroding strut was removed endoscopically with the Nd:YAG laser. Continence has been preserved and ease of catheterization restored.