Karayannacos P E, Shapiro I L, Kakos G S, Chino S, Kilman J W, Vasko J S
Ann Thorac Surg. 1977 Mar;23(3):276-7. doi: 10.1016/s0003-4975(10)64125-2.
This paper presents a previously unrecognized hazard associated with intraaortic balloon pumping: fracture of the catheter due to chemical damage by acetone at the time of removal. Common hospital chemicals such as acetone, ether, and Vi-Drape spray may damage the catheter. Contact between these agents and polyurethane intraaortic balloon catheters should be avoided during dressing changes or repreparation of a sterile operative field. Isopropyl alcohol, Betadine, benzoin, and Cidex do not damage the catheter.