Purified haemoglobin preparations in the evaluation of HbA1c determination by ion exchange chromatography.

作者信息

Cohen M P, Witt J, Wu V Y

机构信息

University City Science Center, Philadelphia, PA 19104.

出版信息

Ann Clin Biochem. 1993 May;30 ( Pt 3):265-71. doi: 10.1177/000456329303000307.

Abstract

The use of ion exchange resins for the estimation of HbA1c in clinical samples rests on the assumption that HbA1c is effectively and efficiently separated from other N-terminally modified haemoglobins and from HbA0. To test this assumption, we applied highly purified preparations of HbA(1a+1b, HbA1c and HbA0 to ion exchange minicolumns, using conditions of application simulating actual blood samples and the first and second elution buffers provided by the manufacturer. The authenticity and purity of the applied haemoglobin preparations were documented by high performance liquid chromatography, gel electrophoresis and carbohydrate content. About 40% of the applied HbA(1a+1b) eluted in the first fraction; 45% eluted in the second fraction, and 10% to 15% required 1 mol/L NaCl to elute from the column. Of the applied HbA1c, 65-80% eluted where expected in the second fraction, about 20% required 1 mol/L NaCl to elute from the column, and the remainder eluted with HbA(1a+1b). Some 3-6% of pure HbA0 applied to minicolumns emerged in the second fraction, with the remainder eluting as expected after making the buffer 1 mol/L in NaCl. The results indicate that the fraction eluting from ion exchange minicolumn chromatography that is designated 'HbA1c' contains HbA(1a+1b), and that a substantial portion of the HbA1c in an applied sample does not elute in this fraction.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验