Zuckier L S, Korupolu G R, Gladshteyn M, Sattenberg R, Goldstein R, Ricciardi R, Goodwin P, Melman A, Blaufox M D
Department of Nuclear Medicine, Albert Einstein College of Medicine, Bronx, New York, USA.
J Nucl Med. 1995 Dec;36(12):2345-51.
We have developed a penile nonimaging scintillation (PNIS) probe consisting of a plastic well-type scintillation crystal interfaced to a portable computer and acquisition board. This report describes the design of the PNIS probe, performance characteristics, mode of usage and illustrative results which demonstrate its capabilities.
With the PNIS probe, penile blood-pool studies were performed in nine patients utilizing 3.7 MBq (100 microCi) autologous 99mTc-labeled red blood cells (RBCs). Venous blood standards were assayed to enable conversion of the count rate to volummetric measurements. Washin of peripherally administered 99mTc-RBCs was mathematically analyzed to estimate penile blood volume and cavernosal flow rate in the flaccid state. The rate of change of penile blood volume after intracavernosal vasodilators was used to generate measures of stimulated flow.
A major advantage of this device over the gamma-camera is a 3300-fold increase in count rate sensitivity, which allows for markedly improved temporal resolution while significantly reducing the radiopharmaceutical dosage. Additionally, the PNIS probe is portable, economical and is not dependent on operator-defined regions of interest. Count rate sensitivity is relatively constant within the bore, with the exception of the proximal region adjacent to the opening, where geometric efficiency is reduced.
The PNIS probe is an effective device for measuring penile activity in radionuclide studies, allowing for acquisition of time-activity curves of the penis during flaccid washin of peripherally labeled red blood cells and after pharmacologic stimulation to induce erection.
我们开发了一种阴茎非成像闪烁(PNIS)探头,它由一个塑料井型闪烁晶体与一台便携式计算机和采集板相连组成。本报告描述了PNIS探头的设计、性能特征、使用方式以及展示其功能的说明性结果。
使用PNIS探头,对9名患者进行阴茎血池研究,使用3.7MBq(100微居里)自体99mTc标记的红细胞(RBC)。测定静脉血标准以将计数率转换为体积测量值。对外周注射的99mTc-RBC的冲洗进行数学分析,以估计疲软状态下的阴茎血容量和海绵体血流率。阴茎海绵体内血管扩张剂注射后阴茎血容量的变化率用于生成刺激血流的测量值。
该设备相对于γ相机的一个主要优点是计数率灵敏度提高了3300倍,这使得时间分辨率显著提高,同时显著降低了放射性药物剂量。此外,PNIS探头便于携带、经济且不依赖于操作员定义的感兴趣区域。除了靠近开口的近端区域几何效率降低外,计数率灵敏度在探头孔内相对恒定。
PNIS探头是放射性核素研究中测量阴茎活性的有效设备,可在周边标记红细胞的疲软冲洗过程中以及药物刺激诱导勃起后获取阴茎的时间-活性曲线。