Barish R J
Cancer Institute, Mary Immaculate Hospital Division, Catholic Medical Center of Brooklyn and Queens, Jamaica, New York 11432, USA.
Med Phys. 1995 Sep;22(9):1467-8. doi: 10.1118/1.597622.
We describe a simplified technique for the acquisition of off-axis beam-quality data (i.e., half-value layers in water) for medical linear accelerators. A measurement protocol is presented in which an ordinary beam-scanning water phantom is used to accurately position an ionization chamber at appropriate measuring positions. Attenuation coefficients are calculated by performing a regression on the measured data. The technique uses the smallest fields for which lateral electron equilibrium can be established.
我们描述了一种用于获取医用直线加速器离轴束流质量数据(即水中半值层)的简化技术。本文介绍了一种测量方案,其中使用普通的束流扫描水模将电离室精确放置在适当的测量位置。通过对测量数据进行回归计算衰减系数。该技术使用能够建立横向电子平衡的最小射野。