Harvey E M, Miller J M, Dobson V
University of Arizona College of Medicine, Department of Ophthalmology, Tucson 85719, USA.
Br J Ophthalmol. 1995 Nov;79(11):983-90. doi: 10.1136/bjo.79.11.983.
To evaluate the overall accuracy and reproducibility of the Alcon portable autokeratometer (PAK) measurements in infants and young children.
The accuracy of the Alcon PAK in measuring toric reference surfaces (1, 3, 5, and 7 D) under various suboptimal measurement conditions was assessed, and the reproducibility of PAK measurements of corneal astigmatism in newborn infants (n = 5), children (n = 19, age 3-5 years), and adults (n = 14) was evaluated.
Measurements of toric reference surfaces indicated (a) no significant effect of distance (17-30 mm) on accuracy of measurements, (b) no systematic relation between amount of toricity and accuracy of measurements, (c) no systematic relation between angle of measurement and accuracy, (d) no difference in accuracy of measurements when the PAK is hand held in comparison with when it is mounted, (e) no difference in accuracy of measurements when axis of toricity is oriented obliquely than when it is oriented horizontally, with respect to the PAK, and (f) a small positive bias (+0.16 D) in measurement of spherical equivalent. The PAK did not prove useful for screening newborns. However, measurements were successfully obtained from 18/19 children and 14/14 adults. There was no significant difference in median measurement deviation (deviation of a subject's five measurements from his/her mean) between children (0.21 D) and adults (0.13 D).
The PAK produces accurate measurements of surface curvature under a variety of suboptimal conditions. Variability of PAK measurements in preschool children is small enough to suggest that it would be useful for screening for corneal astigmatism in young children.
评估爱尔康便携式自动角膜曲率计(PAK)在婴幼儿中的总体准确性和可重复性。
评估了爱尔康PAK在各种非最佳测量条件下测量复曲面参考面(1、3、5和7 D)的准确性,并评估了新生儿(n = 5)、儿童(n = 19,年龄3 - 5岁)和成人(n = 14)角膜散光PAK测量的可重复性。
复曲面参考面的测量表明:(a)距离(17 - 30 mm)对测量准确性无显著影响;(b)散光量与测量准确性之间无系统关系;(c)测量角度与准确性之间无系统关系;(d)PAK手持时与安装时的测量准确性无差异;(e)相对于PAK,复曲面轴倾斜时与水平时的测量准确性无差异;(f)等效球镜测量存在小的正偏差(+0.16 D)。PAK对新生儿筛查无用。然而,成功从18/19名儿童和14/14名成人中获得测量结果。儿童(0.21 D)和成人(0.13 D)的中位测量偏差(受试者五次测量与其平均值的偏差)无显著差异。
PAK在各种非最佳条件下能准确测量表面曲率。学龄前儿童PAK测量的变异性足够小,表明其对幼儿角膜散光筛查有用。