Suppr超能文献

经静脉除颤导联系统的电击后感知性能:心室颤动检测与重新检测分析

Postshock sensing performance in transvenous defibrillation lead systems: analysis of detection and redetection of ventricular fibrillation.

作者信息

Callans D J, Swarna U S, Schwartzman D, Gottlieb C D, Marchlinski F E

机构信息

Clinical Electrophysiology Laboratory, Philadelphia Heart Institute, Presbyterian Medical Center, Pennsylvania 19104, USA.

出版信息

J Cardiovasc Electrophysiol. 1995 Aug;6(8):604-12. doi: 10.1111/j.1540-8167.1995.tb00436.x.

Abstract

INTRODUCTION

The sensing performance of transvenous lead systems may be adversely affected by the delivery of high-energy shocks. This may be due to the proximity of the sensing and energy-delivery electrodes on transvenous leads.

METHODS AND RESULTS

The time required for detection of ventricular fibrillation and redetection after a failed first shock was compared in 93 patients with five different lead system-pulse generator combinations: Cadence--Endotak 60 series, Ventak P--Endotak 60 series, Jewel--Transvene, Cadence--TVL, and Cadence--Transvene. A total of 418 successful and 204 failed first shocks were delivered during induced ventricular fibrillation. Redetection times (RED) were consistently shorter than detection times (DET) in the Jewel-Transvene (RED minus DET: -1.9 +/- 0.8 sec, P < 0.0001), the Cadence-TVL (-1.6 +/- 1.0 sec, P < 0.0001), and the Cadence-Transvene combinations (-2.0 +/- 0.9 sec, P < 0.0004). Redetection times were not significantly different than detection times in the Cadence-Endotak combination (0.9 +/- 3.1 sec; P = 0.09). Redetection times were significantly longer than detection times in the Ventak-Endotak combination (1.2 +/- 2.3 sec; P = 0.034). Prolonged individual redetection episodes (> 8.2 sec) were observed in the Cadence-Endotak (7 [10%] of 73 episodes) and the Ventak-Endotak (4 [10%] of 39 episodes), but not in the Jewel-Transvene, the Cadence-TVL, and the Cadence-Transvene combinations.

CONCLUSIONS

Redetection of ventricular fibrillation may be delayed in some transvenous lead-pulse generator combinations. Successful redetection of ventricular fibrillation following a failed first shock should be demonstrated prior to hospital discharge of patients with implantable defibrillators.

摘要

引言

经静脉导线系统的传感性能可能会受到高能量电击的不利影响。这可能是由于经静脉导线上传感电极和能量输送电极距离较近所致。

方法与结果

对93例患者使用五种不同的导线系统 - 脉冲发生器组合(Cadence - Endotak 60系列、Ventak P - Endotak 60系列、Jewel - Transvene、Cadence - TVL和Cadence - Transvene),比较检测室颤所需时间以及首次电击失败后重新检测的时间。在诱发室颤期间共进行了418次成功的首次电击和204次失败的首次电击。在Jewel - Transvene组合(重新检测时间减去检测时间:-1.9±0.8秒,P < 0.0001)、Cadence - TVL组合(-1.6±1.0秒,P < 0.0001)和Cadence - Transvene组合(-2.0±0.9秒,P < 0.0004)中,重新检测时间始终短于检测时间。在Cadence - Endotak组合中,重新检测时间与检测时间无显著差异(0.9±3.1秒;P = 0.09)。在Ventak - Endotak组合中,重新检测时间显著长于检测时间(1.2±2.3秒;P = 0.034)。在Cadence - Endotak组合(73次发作中的7次[10%])和Ventak - Endotak组合(39次发作中的4次[10%])中观察到个别重新检测时间延长(> 8.2秒),但在Jewel - Transvene、Cadence - TVL和Cadence - Transvene组合中未观察到。

结论

在某些经静脉导线 - 脉冲发生器组合中,室颤的重新检测可能会延迟。对于植入式除颤器患者,在出院前应证明首次电击失败后能成功重新检测到室颤。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验