Felsenberg D, Gowin W, Diessel E, Armbrust S, Mews J
Abt. für Radiologische Diagnostik und Nuklearmedizin, Universitätsklinikum Benjamin Franklin, Freie Universität Berlin, Germany.
Eur J Radiol. 1995 Sep;20(3):179-84. doi: 10.1016/0720-048x(95)00682-g.
The introduction of new devices demands the assessment of their capabilities in established terms. The accuracy, reproducibility and spatial resolution of in vitro (phantom) osteodensitometric and morphometric measurements of QDR 2000 Plus and EXPERT are presented. Design details of these DXA/MXA-devices are listed and discussed in combination with the data acquired in the test measurements and calculations. The image quality will improve with further software developments. The long-term reproducibility and in vivo reliability remains to be evaluated.
新设备的引入需要根据既定标准评估其性能。本文介绍了QDR 2000 Plus和EXPERT体外(体模)骨密度测量及形态测量的准确性、可重复性和空间分辨率。结合测试测量和计算中获取的数据,列出并讨论了这些双能X线吸收仪/混合双能X线吸收仪设备的设计细节。随着软件的进一步开发,图像质量将会提高。长期可重复性和体内可靠性仍有待评估。