Limitations of a pencil beam approach to photon dose calculations in the head and neck region.

作者信息

Hurkmans C, Knöös T, Nilsson P, Svahn-Tapper G, Danielsson H

机构信息

Department of Radiation Physics, University Hospital, Lund, Sweden.

出版信息

Radiother Oncol. 1995 Oct;37(1):74-80. doi: 10.1016/0167-8140(95)01609-k.

Abstract

The inherent limitations of a specific pencil beam model have been studied when applied to a cylindrical geometry simulating the neck region. A comparison is made between measured and calculated absorbed dose in a cylindrical phantom. The goal is to quantify the deviations in the absorbed dose level, i.e., the dose per monitor unit, when photons are used for the treatment of head and neck tumours. Square fields ranging from 5 x 5 up to 30 x 30 cm2 are studied for photon beam energies of 60Co, 4, 6 and 18 MV. Ionisation chamber measurements have been performed in the cylinder as well as in two other configurations in order to trace the origin of possible deviations. For 18 MV no significant deviations are found between measurement and calculation in the cylindrical configuration. For the lower energies, an overestimation of the calculated dose in the cylindrical configuration up to about 6% for a 20 x 20-cm2 60Co field has been found. These deviations have been traced to the basic approximation for the integration volume for phantom scatter calculations inherent in this pencil beam implementation.

摘要

当将特定的笔形束模型应用于模拟颈部区域的圆柱形几何体时,对其固有的局限性进行了研究。在圆柱形体模中对测量的吸收剂量和计算的吸收剂量进行了比较。目的是量化当使用光子治疗头颈部肿瘤时吸收剂量水平(即每个监测单位的剂量)的偏差。针对60Co、4、6和18 MV的光子束能量,研究了面积从5×5到30×30 cm2的方形射野。为了追踪可能偏差的来源,已在圆柱体以及其他两种配置中进行了电离室测量。对于18 MV,在圆柱形配置中测量值与计算值之间未发现显著偏差。对于较低能量,发现在圆柱形配置中计算剂量存在高估,对于20×20 - cm2的60Co射野,高估高达约6%。这些偏差已追溯到该笔形束实现中固有的体模散射计算积分体积的基本近似。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验