Dunseath W J, Kelly E F
Dental Research Center, University of North Carolinia, Chapel Hill 27599-7455 USA.
IEEE Trans Biomed Eng. 1995 Dec;42(12):1212-7. doi: 10.1109/10.476129.
This paper describes a complete, high-performance data-acquisition system for high-resolution EEG. The system includes preamplified EEG electrodes, rigid helmets, differential instrumentation and software-controlled gain amplifiers, digitizing and control circuitry, and optical coupling to a 486 PC. Miniaturized preamplifiers mounted on individual electrodes reduce coupling to external electromagnetic fields and minimize signal distortion caused by increased and/or imbalanced electrode impedances. Electrodes are applied quickly and with minimal skin preparation, yet with high packing densities and repositioning consistent to about 1-3 mm in repeated applications. Extensive testing under realistic conditions demonstrates that the system provides superior electrical performance compared to currently available commercial systems.
本文介绍了一种用于高分辨率脑电图(EEG)的完整、高性能数据采集系统。该系统包括预放大的EEG电极、刚性头盔、差分仪器和软件控制增益放大器、数字化及控制电路,以及与486个人计算机的光耦合。安装在单个电极上的小型化预放大器减少了与外部电磁场的耦合,并将因电极阻抗增加和/或不平衡而导致的信号失真降至最低。电极应用迅速,皮肤准备工作最少,然而具有高组装密度,并且在重复应用中重新定位的误差约为1 - 3毫米。在实际条件下的广泛测试表明,与目前可用的商业系统相比,该系统具有卓越的电气性能。