Influence of the intra-aortic balloon pump on the transcranial Doppler flow pattern in a brain-dead patient.

作者信息

van der Naalt J, Baker A J

机构信息

Critical Care Unit, St Michael's Hospital, Toronto, Canada.

出版信息

Stroke. 1996 Jan;27(1):140-2. doi: 10.1161/01.str.27.1.140.

Abstract

BACKGROUND

Confirmation of clinical brain death with transcranial Doppler (TCD) has been described. With the introduction of mechanical assist devices, it is important to know how these devices influence TCD measurements.

CASE DESCRIPTION

A patient who fulfilled the clinical criteria for brain death after cardiac arrest is described. An intra-aortic balloon pump (IABP) was necessary to maintain hemodynamic stability. A TCD examination was performed as an adjunct to the clinical diagnosis of brain death. A pattern of reversal of blood velocity typical of brain death was observed. With the IABP functioning, an increase of mean forward flow velocity without appreciable increase in the net flow velocities was seen. The results of the TCD measurements with the IABP functioning are not in concordance with values reported in the literature that confirm the clinical diagnosis of brain death.

CONCLUSIONS

Application of TCD in a patient with an IABP could lead to false interpretation of results if the TCD mean velocities are not registered with the IABP on standby or if the net flow velocities are not calculated.