Myocardial tolerance to mechanical actuation is affected by biomaterial characteristics.

作者信息

Anstadt M P, Perez-Tamayo R A, Banit D M, Walthall H P, Cothran R L, Abdel-Aleem S, Anstadt G L, Jones P L, Lowe J E

机构信息

Department of Surgery, Duke University Medical Center, Durham, North Carolina 27710, USA.

出版信息

ASAIO J. 1994 Jul-Sep;40(3):M329-34. doi: 10.1097/00002480-199407000-00018.

Abstract

Direct mechanical ventricular actuation (DMVA) uses a pressure regulated heart cup, fabricated from silicone rubber (SR) for mechanical massage of the heart. Because DMVA has demonstrated potential for long-term circulatory support, investigations are currently exploring the use of more durable materials for fabricating DMVA heart cups. This study assessed the acute effects of heart cups fabricated from SR versus polyurethane (PU) on the myocardium. Dogs (n - 18) received DMVA for 4 hr of ventricular fibrillation (VF) using either SR (n = 10) or PU (n = 8) cups. Microspheres were used to determine perfusion during sinus rhythm (control) and at 2 and 4 hr of support. After support, myocardial biopsies were assayed for high energy phosphate content. Results demonstrated that PU cups required relatively frequent adjustments in drive line parameters that were likely due to material softening during PU cup support. Both PU and SR cups achieved similar hemodynamics during 4 hr of support. Myocardial perfusion, however, demonstrated a marked hyperemia at 4 hr of PU versus SR cup support. Regional high energy phosphate content was significantly decreased in hearts supported by PU versus SR cups. These results suggest that the relatively compliant characteristics of SR materials are important for achieving effective DMVA support without injuring the myocardium.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验