Johnson J E, Boone W R, Blackhurst D W
Greenville Hospital System, South Carolina, USA.
Fertil Steril. 1996 Jan;65(1):156-9. doi: 10.1016/s0015-0282(16)58044-3.
To determine the accuracy and precision of a computer-automated semen analyzer (CASA; Hamilton-Thorne Research, Beverly, MA) over a wide range of sperm concentrations.
Prospective study using comparative measurements of sperm concentration and motility were performed using manual and CASA methods of analyses.
A hospital-based Andrology laboratory.
Male partners of couples undergoing infertility evaluation.
Measurements of sperm concentration in the range of 1 to 200 x 10(6)/mL were performed using hemacytometers and CASA-analyzed 20 MicroCell counting chambers (Conception Technologies, Inc., La Jolla, CA). Accuracy was assessed by comparison to the hemacytometer ("gold standard"). The MicroCell counting chambers also were used for manual and CASA sperm motility determinations.
The CASA-analyzed 20 Microcell Chamber most precisely determined sperm concentration in the 1 to 200 x 10(6)/mL range (intraclass correlation coefficient = 0.94). The hemacytometer was least precise (intraclass correlation coefficient = 0.91). There was a high degree of correlation between the manually and CASA-analyzed MicroCell Chambers in the sperm concentration range of 20 to 149 x 10(6)/mL. Above and below this range, correlation was lower. Motility was determined most precisely using the CASA-analyzed 20 MicroCell Chamber.
The Hamilton-Thorne CASA used with the 20 MicroCell Chamber is a precise and accurate method for determining sperm concentration and motility of semen specimens with concentrations in the 20 to 149 x 10(6)/mL range. This same combination of CASA and MicroCell Chamber provides precise sperm motility results.
确定计算机自动精液分析仪(CASA;汉密尔顿-桑恩研究公司,马萨诸塞州贝弗利)在广泛精子浓度范围内的准确性和精密度。
采用前瞻性研究,使用手动和CASA分析方法对精子浓度和活力进行比较测量。
一家医院的男科实验室。
接受不孕评估夫妇的男性伴侣。
使用血细胞计数器和CASA分析的20微细胞计数室(概念技术公司,加利福尼亚州拉霍亚)对1至200×10⁶/mL范围内的精子浓度进行测量。通过与血细胞计数器(“金标准”)比较评估准确性。微细胞计数室也用于手动和CASA精子活力测定。
CASA分析的20微细胞室在1至200×10⁶/mL范围内最精确地测定精子浓度(组内相关系数=0.94)。血细胞计数器最不精确(组内相关系数=0.91)。在20至149×10⁶/mL的精子浓度范围内,手动和CASA分析的微细胞室之间存在高度相关性。高于和低于此范围,相关性较低。使用CASA分析的20微细胞室最精确地测定活力。
与20微细胞室一起使用的汉密尔顿-桑恩CASA是一种精确且准确的方法,用于测定浓度在20至149×10⁶/mL范围内精液标本的精子浓度和活力。CASA和微细胞室的这种组合提供了精确的精子活力结果。