德国外周定量计算机断层扫描多中心参考数据库。

Multicenter German reference data base for peripheral quantitative computer tomography.

作者信息

Schneider P, Butz S, Allolio B, Börner W, Klein K, Lehmann R, Petermann K, Tysarczyk-Niemeyer G, Wüster C, Zander C

机构信息

Klinik und Poliklinik für Nuklearmedizin, Universität Würzburg, Germany.

出版信息

Technol Health Care. 1995 Oct;3(2):69-73.

PMID:8574764
Abstract

The wide spread use of bone densitometers in Germany and other European countries has required the establishment of a validated reference population data base. A semianthropomorphic forearm cross-calibration phantom (EFP), developed during a concerted research action of the European Union's programme in Biomedical Engineering (COMAC-BME), was used to cross-calibrate the peripheral quantitative computer tomography (pQCT) devices at four German centers participating in the multicenter study. In total, 723 women and 208 men were included in the normal data base. No significant regional differences were found between the data of the different centers. In addition to the manufacturers calibration standard, proper calibration of the pQCT devices could be monitored during collection of the normal female and male data base. As a merit of the COMAC-BME study the measurements obtained with all pQCT devices thus ensured an uniform reference data base for distal radius measurements in Germany.

摘要

骨密度仪在德国和其他欧洲国家的广泛使用要求建立一个经过验证的参考人群数据库。在欧盟生物医学工程计划(COMAC-BME)的一项协同研究行动中开发的半人体前臂交叉校准体模(EFP),被用于对参与多中心研究的四个德国中心的外周定量计算机断层扫描(pQCT)设备进行交叉校准。正常数据库总共纳入了723名女性和208名男性。不同中心的数据之间未发现显著的区域差异。除了制造商的校准标准外,在收集正常女性和男性数据库期间,可以监测pQCT设备的正确校准情况。作为COMAC-BME研究的一个优点,所有pQCT设备获得的测量结果因此确保了德国桡骨远端测量的统一参考数据库。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验