Willmann G, Pfaff H G, Richter H G
Biomed Tech (Berl). 1995 Dec;40(12):342-6. doi: 10.1515/bmte.1995.40.12.342.
Since 1974 Biolox ceramic femoral ball heads have been used successfully for artificial modular hip joints. The revision rate due to ball head fracture is lower than 0.02%. This is an extremely low value. In this article it is shown how the safety of a ceramic ball head can be improved using the procedures of HIPing, engraving by laser technique, and 100% proof testing. By applying these means the materials properties density, grain size, grain size distribution, and the strength of the ball head, i.e. the fracture load can be improved significantly.
自1974年以来,Biolox陶瓷股骨头已成功应用于人工模块化髋关节。因球头骨折导致的翻修率低于0.02%。这是一个极低的值。本文展示了如何通过热等静压处理、激光雕刻技术以及100%验证测试等工艺来提高陶瓷球头的安全性。通过应用这些方法,球头的材料性能如密度、晶粒尺寸、晶粒尺寸分布以及强度(即断裂载荷)可得到显著改善。