Suppr超能文献

Non-invasive monitoring of hemodynamic parameters during hemodialysis.

作者信息

Pizzarelli F, Dattolo P, Piacenti M, Morales M A, Cerrai T, Maggiore Q

机构信息

Department of Nephrology, Ospedale S.M. Annunziata, Firenze, Italy.

出版信息

Int J Artif Organs. 1995 Sep;18(9):499-503.

PMID:8582765
Abstract

We studied in 13 hemodialysis patients intradialytic variations of blood volume (BV) and cardiac output, by means of non-invasive methods. We found a weak correlation, r 0.2 or less, between BV variations and intradialysis blood pressure variations. The sensitivity of the former in describing the variations of the latter was only 32%. During the 30 min preceeding the hypothensive crisis the percent BV variations did not show any predictive trend. On the contrary, refilling increased as blood pressure dropped and a weak inverse relation (r -0.35) was found between these two parameters. Unstable patients had predialytic blood volume values significantly lower than stable ones and comparable to healthy subjects. On the contrary, the correlation between percent variations of cardiac output index and MAP was 0.68 with a sensitivity and specificity of 90% and 59%, respectively. Unfortunately these promising results were obtained only with an estimate of cardiac output obtained by echocardiography and not by transthoracic impedance cardiography, which is much more feasible than the former as on-line monitoring of cardiac output. On-line monitoring of hemodynamic parameters is an appealing but still unsolved task.