Becker T, Schmidt H L, Lechner E
Lehrstuhl für Allgemeine Chemie und Biochemie, Technische Universität München, Freising-Weihenstephan, Germany.
Z Lebensm Unters Forsch. 1995 Dec;201(6):537-40. doi: 10.1007/BF01201579.
A flow-injection system is presented, which is versatile, automated and has an integrated enzyme system. Its efficiency at determining D- and L-lactate separately in dairy products is compared with the officially used reference method (DIN 10335). By means of a mathematical elimination of matrix effects, the pretreatment steps can be clearly reduced. As a result of the immobilization of the enzymes, an appropriate automation software and an adaptation of the system developed to the given medium, the analysis could be performed within 5 min and was distinctly cheaper than alternative methods. The values determined in 15 different dairy products deviated from those of the reference method by 3% at a maximum.
介绍了一种流动注射系统,该系统用途广泛、自动化且具有集成酶系统。将其在乳制品中分别测定D-乳酸和L-乳酸的效率与官方使用的参考方法(DIN 10335)进行了比较。通过数学方法消除基质效应,可以显著减少预处理步骤。由于酶的固定化、合适的自动化软件以及对所开发系统进行了针对给定介质的调整,分析可在5分钟内完成,且明显比其他方法成本更低。在15种不同乳制品中测定的值与参考方法的值相比,最大偏差为3%。