Brodkey M J, Schlegel J U, Derouen T A
Invest Urol. 1977 May;14(6):417-20.
An experimental technique has been devised to determine effective renal plasma flow, using the gamma scintillation camera. This involves a computer analysis of the first 3 min of the renal histogram and does not require serum or urine collections. This technique is not sufficiently accurate to provide an exact value of plasma flow, but it can be used to provide an approximate value and hence used as a screening tool.
已设计出一种使用γ闪烁照相机来测定有效肾血浆流量的实验技术。这涉及对肾直方图前3分钟进行计算机分析,且不需要采集血清或尿液。该技术的准确性不足以提供血浆流量的确切值,但可用于提供一个近似值,因此可用作一种筛查工具。