The purpose of this study was to compare hydroxyapatite (HA)-coated titanium (Ti) and HA-coated cobalt-chrome (CoCr) implants in the distal femur of the rabbit by evaluating bone apposition and interfacial shear strength. Bilateral cylindrical implants with a plasma sprayed 50-microns thick HA coating were press-fit into the metaphyseal cancellous bone of the lateral femoral condyles in a transverse fashion, and the animals were sacrificed at 2, 4, and 8 weeks postimplantation. Mechanical strength of the interface between HA and bone was measured using the pushout method. For histologic analysis, the fractional linear extent of bone apposition was quantitated. No differences were found in the interfacial shear strength between the Ti and CoCr at any time period. The amount of bone apposition increased significantly at each time interval for both substrate metals, but there were no significant differences between the two substrates at any of the time periods studied. The HA-coated CoCr implants performed in a similar manner to the HA-coated Ti implants, both mechanically and histologically, suggesting that HA-coated CoCr implants deserve further study as a viable alternative to Ti for the biological fixation of total joint components in orthopaedic surgery.
本研究的目的是通过评估骨附着和界面剪切强度,比较羟基磷灰石(HA)涂层钛(Ti)和HA涂层钴铬(CoCr)植入物在兔股骨远端的情况。将具有50微米厚等离子喷涂HA涂层的双侧圆柱形植入物以横向方式压配合到股骨外侧髁的干骺端松质骨中,并在植入后2、4和8周处死动物。使用推出法测量HA与骨之间界面的机械强度。对于组织学分析,对骨附着的线性分数范围进行定量。在任何时间段,Ti和CoCr之间的界面剪切强度均未发现差异。两种基底金属在每个时间间隔的骨附着量均显著增加,但在研究的任何时间段,两种基底之间均无显著差异。HA涂层CoCr植入物在机械和组织学方面的表现与HA涂层Ti植入物相似,这表明HA涂层CoCr植入物作为骨科手术中全关节组件生物固定的Ti的可行替代品值得进一步研究。