Cook S D, Thomas K A, Kay J F, Jarcho M
Department of Orthopaedic Surgery, Tulane University School of Medicine, New Orleans, Louisiana 70112.
Clin Orthop Relat Res. 1988 Jul(232):225-43.
The interface mechanical characteristics and histology of commercially pure (CP) titanium- and hydroxyapatite- (HA) coated Ti-6Al-4V alloy were investigated. Interface shear strength was determined using a transcortical push-out model in dogs after periods of three, five, six, ten, and 32 weeks. Undecalcified histologic techniques with implants in situ were used to interpret differences in mechanical response. The HA-coated titanium alloy implants developed five to seven times the mean interface strength of the uncoated, beadblasted CP titanium implants. The mean values for interface shear strength increased up to 7.27 megaPascals (MPa) for the HA-coated implants after ten weeks of implantation, and the maximum mean value of interface shear strength for the uncoated CP titanium implants was 1.54 MPa. For both implant types there was a slight decrease in mean shear strength from the maximum value to that obtained after the longest implantation period (32 weeks). Histologic evaluations in all cases revealed mineralization of interface bone directly onto the HA-coated implant surface, with no fibrous tissue layer interposed between the bone and HA visible at the light microscopic level. The uncoated titanium implants had projections of bone to the implant surface with apparent direct bone-implant apposition observed in some locations. Measurements of the HA coating material made from histologic sections showed no evidence of significant HA resorption in vivo after periods of up to 32 weeks.
研究了商业纯钛(CP钛)及羟基磷灰石(HA)涂层的Ti-6Al-4V合金的界面力学特性和组织学。采用经皮质推出模型,在犬体内植入后3周、5周、6周、10周和32周时测定界面剪切强度。采用未脱钙组织学技术对植入原位的植入物进行研究,以解释力学反应的差异。HA涂层钛合金植入物的平均界面强度是未涂层、经喷砂处理的CP钛植入物的5至7倍。植入10周后,HA涂层植入物的界面剪切强度平均值增至7.27兆帕(MPa),未涂层CP钛植入物的界面剪切强度最大值为1.54 MPa。两种植入物类型的平均剪切强度均从最大值到最长植入期(32周)后的数值略有下降。所有病例的组织学评估均显示,界面骨直接矿化在HA涂层植入物表面,在光学显微镜下,骨与HA之间未见纤维组织层。未涂层钛植入物有骨向植入物表面的突出,在某些部位观察到明显的骨与植入物直接贴合。从组织学切片测量HA涂层材料发现,在长达32周的时间里,体内未出现明显的HA吸收迹象。