Suppr超能文献

乳腺摄影体模材料的组成。

Composition of mammographic phantom materials.

作者信息

Geise R A, Palchevsky A

机构信息

Department of Radiology, University of Minnesota Hospital and Clinics, Minneapolis 55455, USA.

出版信息

Radiology. 1996 Feb;198(2):347-50. doi: 10.1148/radiology.198.2.8596830.

Abstract

PURPOSE

To determine the composition of a breast phantom that is most suitable for phototimer testing.

MATERIALS AND METHODS

Kilovolt peak and milliampere second values were analyzed for phototimed exposures of phantoms containing 0%, 30%, 50%, 70%, and 100% simulated glandular material at thicknesses of 2, 4, 6, and 8 cm. Data also were analyzed from 1,578 mammograms in 417 patients. The mean effective glandular content of phantoms that simulated patient breasts of varying thicknesses was determined.

RESULTS

Examination of mammographic techniques showed that the effective glandular content that simulated the average patient breast was 34%. Effective glandular content was 16% in breasts at least 7 cm thick when compressed; 26%, more than 5 to up to 7 cm thick; 42%, more than 3 to up to 5 cm thick, and 68% in breasts at most 3 cm thick.

CONCLUSION

A phantom composed of 30% glandular tissue and 70% adipose tissue allows closer simulation of the phototimer response of the mammographic x-ray unit for the average breast. The phantom currently used contains 16% more glandular tissue than the average breast.

摘要

目的

确定最适合用于光定时器测试的乳腺体模的组成。

材料与方法

分析了千伏峰值和毫安秒值,这些值来自于含有0%、30%、50%、70%和100%模拟腺体材料的体模在2厘米、4厘米、6厘米和8厘米厚度下的光定时曝光。还分析了417例患者的1578张乳房X线照片的数据。确定了模拟不同厚度患者乳房的体模的平均有效腺体含量。

结果

对乳房X线摄影技术的检查表明,模拟平均患者乳房的有效腺体含量为34%。压缩后至少7厘米厚的乳房中有效腺体含量为16%;5至7厘米厚的乳房中为26%;3至5厘米厚的乳房中为42%;最多3厘米厚的乳房中为68%。

结论

由30%腺体组织和70%脂肪组织组成的体模能够更接近地模拟平均乳房的乳房X线摄影X射线装置的光定时器响应。目前使用的体模比平均乳房多含16%的腺体组织。