A dual sensitivity screen system for TMJ image enhancement in cephalometric radiography: sensitometric evaluation.

作者信息

Wakoh M, Farman A G, Nishikawa K, Kuroyanagi K, Scarfe W C, Braun S

机构信息

Division of Radiology and Imaging Sciences, School of Dentistry, University of Louisville, KY, USA.

出版信息

Dentomaxillofac Radiol. 1995 Aug;24(3):191-4. doi: 10.1259/dmfr.24.3.8617394.

Abstract

OBJECTIVES

An in vitro evaluation of the image quality of the TMJ OrthoCeph Slimline Cassette System (TOSCS).

METHODS

The density response, resolution and clarity of the system were evaluated by use of the characteristic curve, modulation transfer function and root mean square quantum mottle. Relative sharpness was evaluated qualitatively by five observers.

RESULTS

Increased density with Trimax 12 gave almost the same contrast as that with Trimax 8 in the diagnostic range, whereas MTF and RMS for Trimax 12 were inferior.

CONCLUSIONS

The dual-sensitivity screen system should lead to improved radiographic contrast of the TMJ region in cephalometric radiography. The reduction in sharpness is probably clinically insignificant.