Suppr超能文献

无牙颌中种植体支持修复体适合度精度的体内测量。

In vivo measurements of precision of fit involving implant-supported prostheses in the edentulous jaw.

作者信息

Jemt T

机构信息

Brånemark Clinic, Public Dental Health, Göteborg, Sweden.

出版信息

Int J Oral Maxillofac Implants. 1996 Mar-Apr;11(2):151-8.

PMID:8666445
Abstract

The objective of this study was to measure and compare the precision of fit of implant-supported prostheses in the edentulous jaw by using both the master cast replicas and the intraoral implants as references. Seven maxillary and 10 mandibular prostheses were randomly selected and measured by means of a three-dimensional (3-D) photogrammetric technique. The results indicated that prostheses routinely connected to osseointegrated implants could demonstrate distortion between the framework and individual implants of up to several hundred microns. When master casts were used as a reference, the mean 3-D distortion of the center point of gold cylinders was 37 microns (SD 18) and 75 microns (SD 40) for mandibular and maxillary prostheses, respectively. The corresponding mean displacement was 90 microns (SD 51) and 111 microns (SD 59), respectively, when the intraoral implants were used as references. The mean 3-D distortion was significantly higher for the intraoral measurements in both arches (P < .001 and P < .05). Furthermore, the overall distortion was significantly higher for the maxillary prostheses when the master casts were used as the reference (P < .05). However, for the intraoral measurements, no statistically significant difference of fit between the arches was possible to observe (P < .05). This could possibly be explained by the finding that intraoral measurements of the mandibular prostheses indicated a deformation and rotation of the mandible that was not observed in the maxillary prostheses. A further factor in the lack of statistical significance could be the relatively small sample size.

摘要

本研究的目的是通过使用主模型复制品和口内种植体作为参考,测量并比较无牙颌中种植体支持修复体的适合精度。随机选择7个上颌和10个下颌修复体,采用三维(3-D)摄影测量技术进行测量。结果表明,常规连接至骨整合种植体的修复体在支架与单个种植体之间可表现出高达数百微米的变形。以主模型作为参考时,下颌和上颌修复体中金柱中心点的平均三维变形分别为37微米(标准差18)和75微米(标准差40)。以口内种植体作为参考时,相应的平均位移分别为90微米(标准差51)和111微米(标准差59)。两个牙弓的口内测量中平均三维变形均显著更高(P <.001和P <.05)。此外,以主模型作为参考时,上颌修复体的整体变形显著更高(P <.05)。然而,对于口内测量,两个牙弓之间的适合度无统计学显著差异(P <.05)。这可能是由于以下发现所解释:下颌修复体的口内测量显示下颌骨有变形和旋转,而上颌修复体中未观察到。缺乏统计学显著性的另一个因素可能是样本量相对较小。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验