Russell S H, Freeman J W
Department of Anaesthesia and Intensive Care, Queen Elizabeth Hospital, Birmingham.
Anaesthesia. 1996 Apr;51(4):338-40. doi: 10.1111/j.1365-2044.1996.tb07743.x.
In this study we compared urinary bladder and oesophageal temperatures with the core temperature obtained from a pulmonary artery catheter in patients undergoing orthotopic liver transplantation. The bladder temperature was a closer approximation to pulmonary artery temperature in this group of patients than the oesophageal temperature and has much to recommend it in terms of convenience, safety and postoperative patient comfort.
在本研究中,我们比较了接受原位肝移植患者的膀胱温度和食管温度与通过肺动脉导管测得的核心温度。在这组患者中,膀胱温度比食管温度更接近肺动脉温度,并且在便利性、安全性和术后患者舒适度方面有诸多优势。