Jornet N, Ribas M, Eudaldo T
Servei de Radiofisica i Radioprotecció, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain.
Radiother Oncol. 1996 Mar;38(3):247-51. doi: 10.1016/0167-8140(95)01677-5.
The aim of this paper is to discuss the measurements carried out to implement 'in vivo dosimetry' with EDP-30 diodes in total body irradiation (TBI) techniques. Exit calibrations and calibrations behind cerrobend protection blocks showed the importance of calibrating diodes in all relevant clinical conditions. Special attention was given to calibration of diodes behind cerrobend blocks. Dependence of the calibration factors on the thickness of the shielding blocks was, therefore, studied. This dependence was again studied after adding a wax cap to the diode and when the ionisation chamber was placed at the same depth as the measuring point of the diode. Temperature dependence in diode sensitivity and dependence on accumulated dose for diodes response and for temperature correction factors were also examined.
本文旨在讨论在全身照射(TBI)技术中使用EDP-30二极管进行“体内剂量测定”时所开展的测量工作。出射校准以及在铈镧合金保护块后的校准表明了在所有相关临床条件下对二极管进行校准的重要性。特别关注了铈镧合金块后二极管的校准。因此,研究了校准因子对屏蔽块厚度的依赖性。在给二极管添加蜡帽后以及将电离室放置在与二极管测量点相同深度时,再次研究了这种依赖性。还检查了二极管灵敏度的温度依赖性以及二极管响应和温度校正因子对累积剂量的依赖性。