Transesophageal echocardiography in fetal sheep. A monitoring tool for open and fetoscopic cardiac procedures.

作者信息

Kohl T, Stelnicki E J, VanderWall K J, Szabo Z, Ko E, Bruch S W, Harrison M R, Silverman N H, Hanley F L, Chou T M

机构信息

Division of Pediatric Cardiothoracic Surgery, University of California, San Francisco, 505 Parnassus Avenue, San Francisco, CA 94143, USA.

出版信息

Surg Endosc. 1996 Aug;10(8):820-4. doi: 10.1007/s004649900169.

Abstract

BACKGROUND

Cardiac procedures in exteriorized fetuses or assisted by fetoscopy require monitoring capabilities not attended by conventional maternal transabdominal echocardiography.

METHODS

We, therefore, assessed the potential of fetal transesophageal echocardiography (TEE) utilizing an intravascular ultrasound catheter (IVUC) for fetal cardiac monitoring. We inserted a 10-F-10-MHz IVUC into the esophagus in 12 exteriorized fetal sheep and by a fetoscopic approach in 4 fetal sheep. Cardiac events were observed. Heart rate, cardiac rhythm, patency of the foramen ovale and ductus arteriosus, and the width of the branch pulmonary arteries could be assessed in all fetuses. Ventricular contractility could be assessed only in fetuses weighing less than 2.5 kg. Larger fetuses did not allow adequate imaging of the apical portion of the ventricles because of limited tissue penetration of the IVUC. Fetal TEE permitted placing small guide wires in the cardiac atria and left ventricle. Short-lived premature beats following intracardiac manipulations of these wires could be observed by fetal TEE in all cases.

RESULTS

At autopsy, no complications from IVUC insertion were observed in the exteriorized fetuses. Fetoscopic placement of the IVUC resulted in minor perioral skin erosion in two nonexteriorized fetuses.

CONCLUSIONS

In conclusion, fetal TEE can be achieved with minor fetal injury and may provide useful information during open and fetoscopic cardiac procedures. Further improvements in IVUC design will permit the application of this technique to monitor human fetal cardiac procedures.

摘要

背景

对体外胎儿进行心脏手术或在胎儿镜辅助下进行心脏手术需要常规经腹超声心动图所不具备的监测能力。

方法

因此,我们评估了利用血管内超声导管(IVUC)进行胎儿经食管超声心动图(TEE)用于胎儿心脏监测的潜力。我们将一根10F-10MHz的IVUC经食管插入12只体外胎儿绵羊体内,并通过胎儿镜方法插入4只胎儿绵羊体内。观察心脏事件。所有胎儿均可评估心率、心律、卵圆孔和动脉导管的通畅情况以及肺分支动脉的宽度。仅在体重小于2.5kg的胎儿中可评估心室收缩力。由于IVUC的组织穿透性有限,较大的胎儿无法对心室顶端部分进行充分成像。胎儿TEE可在心脏心房和左心室内放置小导丝。通过胎儿TEE在所有病例中均可观察到对这些导丝进行心内操作后出现的短暂性早搏。

结果

尸检时,在体外胎儿中未观察到IVUC插入引起的并发症。通过胎儿镜放置IVUC导致2只非体外胎儿出现轻微的口周皮肤糜烂。

结论

总之,可以在对胎儿造成轻微损伤的情况下实现胎儿TEE,并且在开放性和胎儿镜心脏手术期间可能提供有用信息。IVUC设计的进一步改进将使该技术能够应用于监测人类胎儿心脏手术。