Cooley D A
Med Instrum. 1977 Mar-Apr;11(2):82-4.
Thirty years have elapsed since the introduction of the first artifical heart valves. Since that time, many improvements have been made in the design and fabrication of prosthethic valves, but the perfect valve remains elusive. After experience with clinical implantation of more than 5000 artificial valves of all types, we designed a new prosthesis which we believe incorporates the best features of the earlier models and reflects our concepts of the best configuration of an artificial valve. Thus far we have implanted over 2500 Cooley-Cutter valves and the results have been gratifying. The valve has a full-orifice design with the metallic frame milled from a single block of medical-grade titanium, has a pyrolite occluder in a double-cone configuration, and Teflon sewing rings to conform to the aortic and mitral annuli. A major problem with prostheses made from synthetic materials is the need for prolonged anticoagulation. The incidence of thromboembolic complications with this valve has been very low and excellent late results encourage us to use this prosthesis routinely except in unusual circumstances.
自首个人工心脏瓣膜问世以来,三十年已经过去了。从那时起,人工瓣膜的设计和制造有了许多改进,但完美的瓣膜仍然难以实现。在临床植入了超过5000个各种类型的人工瓣膜之后,我们设计了一种新的假体,我们认为它融合了早期模型的最佳特性,并反映了我们对人工瓣膜最佳构造的理念。到目前为止,我们已经植入了超过2500个库利-卡特瓣膜,结果令人满意。该瓣膜采用全开口设计,金属框架由一整块医用级钛铣削而成,有一个双锥构型的热解碳封堵器,以及用于贴合主动脉和二尖瓣瓣环的特氟龙缝合环。由合成材料制成的假体的一个主要问题是需要长期抗凝。这种瓣膜的血栓栓塞并发症发生率非常低,出色的后期效果鼓励我们在一般情况下常规使用这种假体。