Resano F G, Goldstein S A, Boyce S W
Department of Cardiac Anesthesia, Washington Hospital Center, Washington, DC, USA.
ASAIO J. 1996 May-Jun;42(3):240-1.
A case of systemic embolization arising from an akinetic left ventricle in a peripartum cardiomyopathy patient supported with the Abiomed BVS-5000 ventricular assist device is reported. In an akinetic ventricle, prevention of blood stasis by positioning the atrial cannula tip across the mitral valve to decompress the ventricle may prevent this complication.
报道了一例围产期心肌病患者,其左心室运动不能,在使用Abiomed BVS - 5000心室辅助装置支持下发生全身栓塞的病例。在运动不能的心室中,将心房插管尖端置于二尖瓣上方以减压心室,预防血液淤滞,可能会预防这种并发症。