Suppr超能文献

Hepatic artery: variability in measurement of resistive index and systolic acceleration time in healthy volunteers.

作者信息

Paulson E K, Kliewer M A, Frederick M G, Keogan M T, Delong D M, Nelson R C

机构信息

Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA.

出版信息

Radiology. 1996 Sep;200(3):725-9. doi: 10.1148/radiology.200.3.8756922.

Abstract

PURPOSE

To determine the intrinsic sampling variability of measurements of hepatic artery resistive index and systolic acceleration time in healthy subjects and to estimate the components of variability attributable to the sonographer, individual measurement, and subject.

MATERIALS AND METHODS

In a randomized, controlled (blinded) fashion, nine sonographers measured hepatic artery resistive index and systolic acceleration time in five healthy subjects by using Doppler ultrasound (US). Analysis of variance was used to estimate the contribution of several factors to the observed variability in measurements.

RESULTS

The standard deviation for a single measurement was 0.08 for resistive index and 39 msec for systolic acceleration time. For resistive index, the estimated variance components were 0.0012 (18%), 0.0004 (6%), and 0.0050 (76%) for the subject, sonographer, and intrinsic variability, respectively. For systolic acceleration time, the estimated variance components were 59 msec (4%), 264 msec (17%), and 1,250 msec (79%) for the subject, sonographer, and intrinsic variability, respectively.

CONCLUSION

Because of substantial variability in hepatic arterial measures, caution is indicated when interpreting small changes in the measurement of these Doppler US indexes.