Biggs D S, Thomson E S
Department of Medical Physics and Bioengineering, Norfolk and Norwich Health Care NHS Trust, UK.
Br J Radiol. 1996 Jun;69(822):544-7. doi: 10.1259/0007-1285-69-822-544.
An alternative technology for performing interstitial intracranial radiosurgery is presented. Treatments are performed intraoperatively using a stereotactically mounted, miniature X-ray device. A concentrated, spherical pattern of X-rays is generated about the tip of a needle-like probe, which is inserted into the site of the lesion allowing the absorbed dose to healthy tissue to be minimized. A description of the device is given and results of investigations into the operating characteristics discussed.